Pulse rate measurement
Power supply
External DC power supply (isolated)
Internal battery
Housing construction
Environmental conditions
Electromagnetic compatibility
Potential impact of electromagnetic
disturbances
Air filter
Air tubing
IEC 60601-1 classifications
Range: 40 bpm to 240 bpm, Resolution: 1 bpm, Accuracy: Refer to Nonin Xpod technical
specifications at www.nonin.com.
AC 100–240V, 50–60Hz, 2.2 A, max. 65 W
24 V, 3 A
Lithium-Ion battery, 14.4 V, 2.75 Ah, 40 Wh
Operating hours: 3 h with a new battery under normal conditions (see below).
Patient type: home chronic; pressure: IPAP/EPAP 15/5 cm H
tubing: 2 m; leak: 0; respiratory rate: 20 bpm; battery capacity: 100%
Patient type: hospital acute; pressure: IPAP/EPAP 20/5 cm H
tubing: 2 m; leak: 0; respiratory rate: 45 bpm; battery capacity: 100%
Flame retardant engineering thermoplastic
•
Operating temperature: +32°F (0°C) to + 95°F (35°C)
•
Operating humidity: 10%–95% non-condensing
•
Storage and transport temperature: -4°F (-20°C) to +140°F (60°C); +122°F (+50°C*)
•
Storage and transport humidity: 10%–95% non-condensing
•
Air pressure: 680hPa to 1,100hPa; Altitude: 9,842' 5" (3,000 m)
Note: The performance may be limited below 800 hPa.
*NONIN XPOD
The device complies with all applicable electromagnetic disturbances requirements (EMD)
according to IEC 60601-1-2:2014, for residential, commercial and light industry
environments.
It is recommended that mobile communication devices are kept at least 0.3 m away from
the device.
Information regarding the electromagnetic emissions and immunity of this ResMed device
can be found in www.resmed.com/downloads/devices.
The loss or degradation of the following clinical functions due to electromagnetic
disturbances could result in compromised patient safety:
•
Accuracy of ventilation control
•
Accuracy of monitoring of airway pressure and volume
•
Therapy alarms
Detection of this degradation could be observed by the following device behaviour:
•
Erratic ventilation delivery
•
Rapid fluctuations in monitored parameters
•
False activation of therapy or technical alarms
Electrostatic fibre mesh with TPE frame structure. Bacterial filtration efficiency of >99.9%
for ~2.7 micron particle size.
Standard air tubing: Flexible plastics, 6'6" (2 m) or 9'10" (3 m) length (0.9"/22 mm diameter)
SlimLine air tubing: Flexible plastics, 6' (1.83 m) length (0.6"/15 mm diameter)
For more information on air tubing resistance to flow, compliance, tubing length, and inner
diameter, refer to the Air tubing compliance guide in ResMed.com.
•
Class II (Clause 3.14–double insulation). This adherence means the need for an
protective earthing (ie, an earthed plug) is not necessary.
•
Type BF
•
Continuous operation
O; mask type: Ultra Mirage; air
2
O; mask type: Ultra Mirage; air
2