Figure 4
1
Keyboard push-button for LOW SPEED mode selection. It is active only when the front
panel operation has been selected. Pressed once, the pump runs at about 2/3 of the
nominal speed. To unselect the mode, press the push-button again.
2
Keyboard push-button for START, STOP, RESET mode selection. It is active only when the
front panel operation has been selected. By pressing once the starting phase begins; if
pressed again it stops the pump. If the pump has been stopped automatically by a fault,
this push-button must be pressed once to reset the controller and a second time to restart
the pump.
3
Keyboard push-button to recall on the display the cycle number, cycle time and pumplife.
4
Keyboard push-button to recall on the display the pump current, pump temperature, pump
power and rotational speed. It is always active regardless of the operating mode selected.
Push-buttons 3 and 4, if pressed together for at least 2 seconds put the controller in a
routine where it is possible to program some operation parameters.
5
LCD back-lighted alphanumeric display: dot matrix 2 lines x 16 characters.
If a controller is to be scrapped, it must be disposed of in accordance
with the specific national standards.
Turbo-V 1000HT PCB Controller User Manual / 87-900-906-01