FOREFOOT REMOVAL
(ACCESS TO FRONT BUMPER AND LOWER BUSHINGS)
1.
To remove the PreLoad Insert, position the PreLoader Tool on the foot to remove the load off of the
insert. Tighten the handle down on the forefoot then swing the cam lever to either side.
For detailed instructions refer to the Preloader Tool Instruction Sheet.
2.
Remove the front bumper and axial pin as instructed (above) for the Ankle Bone.
3.
Remove the forefoot bone to access the lower bushings.
4.
For re-assembly, lubricate the axial pin, inside and outside of the lower bushings, then reverse steps 1-4.
Torque the axial pin screw to 4 N·m (36 in·lbs)
GAIT MATCHING
Contact College Park Technical Service if you have questions about modifying the gait match.
WARNING
Do not expose this product to pH extremes or corrosive materials (water, salt water or other liquids).
▪
Failure to follow these technical instructions or use of this product outside the scope of its Limited
▪
Warranty may result in injury to the patient or damage to the product.
Disassembly, modification or removal from the foot shell, or any further disassembly or modification of
▪
components will void the warranty.
WARRANTY INSPECTION / MAINTENANCE INFORMATION
College Park recommends that you schedule your patients for check-ups according to the
warranty Inspection schedule below.
High patient weight and/or impact level may require more frequent inspections. Soft component wear depends on
the patient weight, impact level and environment. We recommend you inspect the following applicable parts for
excessive wear and fatigue at each warranty inspection and replace as needed.
▪ Soft components (disassemble, inspect & re-lubricate)
▪ Composites and Adapters
WARRANTY INSPECTION SCHEDULE FOR TRUSTEP: SIX MONTHS, THEN ANNUALLY.
TECHNICAL ASSISTANCE /
EMERGENCY SERVICE 24-7-365
College Park's regular office hours are Monday through Friday, 8:30 am – 5:30 pm (EST).
After hours, an emergency Technical Service number is available to contact a College
Park representative.
LIABILITY
The manufacturer is not liable for damage caused by component combinations that were not authorized by the
manufacturer.
CAUTION
College Park products and components are designed and tested according to the applicable official standards
or an in-house defined standard when no official standard applies. Compatibility and compliance with these
standards are achieved only when College Park products are used with other recommended College Park
components. This product has been designed and tested based on single patient usage. This device should NOT
be used by multiple patients. If any problems occur with the use of this product, immediately contact your
medical professional.