b. The touch screen height may be adjusted by releasing the locking lever on the main post
and pressing the release button on the top of main body pole as illustrated in Figure 8.
The touch screen tilt may be adjusted by loosening the lock behind the touch screen and
moving the screen to the desired position as illustrated in Figure 9.
c. The adjustable tray can accommodate two Drivers as well as Driver cables and Probe
tubing. The position of the tray may be adjusted by releasing the locks on the tray and
rotating, lifting, or lowering the tray to the desired height as illustrated in Figure 10.
Figure 7 – Locking the system casters
Figure 9 – Adjusting the touch screen tilt
E
®
3.
Turn on the main power via the Power Switch on the rear of the console as illustrated in Figure 11.
Figure 11 – Main Power Switch
4.
Turn on the standby power via the switch on the front of the system as illustrated in Figure 12.
Figure 12 – Standby Power Switch
in Figure 13. The active Driver is connected by a green LED light above the connector. Reference
the applicable Driver Instructions for Use for further information on Driver setup. If desired,
connect the appropriate Foot Pedal to the connector on the rear of the console as illustrated in
Figure 14.
Figure 8 – Adjusting the touch
screen height
Figure 10 – Adjusting the tray position
Figure 13 – Driver Connectors
Figure 14 – Foot pedal connector
6.
Follow the touch screen prompts. Install and calibrate the Biopsy Probe. Reference the applicable
Probe Instructions for Use for further information on Probe setup.
Operating the System via the Touch Screen
for Use.
8.
The
®
Breast Biopsy System status is displayed on the touch screen as shown in
Figure 15. The sample acquisition options may be selected via this screen.
the touch screen illustrated in Figure 15.
screen as illustrated in Figure 15
full sampling.
button on the touch screen as illustrated in Figure 15.
return to normal tissue sampling.
Half (Sample button)
Dense (Tissue button)
9.
During the procedure, the
to apply a saline rinse to samples collected in the sample trap. Initiate the rinse by pressing the
"Rinse" button on the touch screen.
10.
If needed, the
®
Breast Biopsy System allows the operator to administer
additional anesthesia through the
anesthetic option to sweep the sample notch through the desired pattern while administering
the Foot Pedal.
Figure 15 – The touch screen – Sample Tab
Full (Sample button)
Normal (Tissue button)
®
Breast Biopsy System has an optional Rinse Mode
Rinse (button)
®
Figures 15 and 16. Select the desired