EN
Instructions
NOTES:
• Recharge battery packs before each surgery, even if they
have not been used, to ensure the maximum running time.
• The charge cycle takes approximately 5 to 60 minutes.
• The expected run time for each battery is 2.5 to
10 minutes depending on demand procedure utilization.
• Only use battery packs that have been charged within the
last 48 hours.
• Store battery packs on the charger where they will be
maintained fully charged until they are sterilized. DO
NOT store battery packs in handpieces where they
will discharge. Discharge will occur even though the
handpiece is not running.
• Maintenance documentation for this equipment is available
upon request to Stryker-authorized service personnel only.
1. To properly install the battery pack into a handpiece,
read and understand the appropriate Stryker handpiece
instructions for use.
2. To properly charge the battery pack, read and understand
the instructions for use provided with the Stryker System 6
Battery Charger (REF 6110-120-000).
4
Care Instructions
For processing instructions and disposal/recycle information,
see the care instructions manual supplied with the handpiece.
6215-001-700 Rev-AA
www.stryker.com