Specifications
Display:
DM-100: LCD (3200) and 70-segment bar graph
In autorange mode, ranges change at approximately 3200 counts. In manual range
mode, range extends to 3400 counts.
DM-110: LCD (4000) and 82-segment bar graph
Polarity: Automatic
Sampling Rate:
Numeric Display: 2 per second
Bar Graph Display: 12 per second
Temperature Coefficient: 0.15 x (Accuracy) per °C below 18 °C or above 28 °C
Automatic Power-Off:
DM-100: After 10 minutes of inactivity
DM-110: After 30 minutes of inactivity
Overload Protections:
Volts: See Overvoltage Protection Categories
Amps: A: 16A/500V HBC F Fuse, Interrupting Rating 10 kA, 1/4 x 1-1/4"
Other Functions: 600 VDC/VAC RMS
Overvoltage Protection Categories: VΩµA Terminal:
Category III, 600 VAC and 600 VDC
Category II, 1000 VAC and 1000 VDC
Operating Conditions:
0 °C to 30 °C (32 °F to 86 °F), 0% to 80% Relative Humidity
30 °C to 40 °C (86 °F to 104 °F), 0% to 75% Relative Humidity
40 °C to 50 °C (104 °F to 122 °F), 0% to 45% Relative Humidity
Indoor use only.
Storage Conditions:
-20 °C to 60 °C (-4 °F to 140 °F), 0% to 80% Relative Humidity
Remove batteries
Elevation: 2000 m (6500') maximum
Pollution Degree: 2
Battery:
DM-100: Two AAA 1.5-Volt batteries (NEDA 24A or IEC LR03)
DM-110: One 9-Volt battery (NEDA 1604 or IEC 6F22)
Battery Life:
DM-100: Approximately 500 hours with alkaline
DM-110: Approximately 300 hours with alkaline