Chapter
6

Acquisition of Radiographic Images

Turn on the PC that has the GXS-700 installed and launch the imaging software, such as
VixWin Platinum (please refer to the software manual for information on the software
program).
1.
Carefully plug the USB end of the sensor cable into the PC USB port. Do not use force,
ensure that the plug is properly oriented with the PC port. Try rotating the USB plug
if it does not easily insert into port.
2.
Set the required technical settings (exposure time, etc.) on the X-ray generator (refer
to X-ray Dose / Exposure Time Settings on page 6-7.
3.
Assemble the rings, bars and holders so that the entire sensor area is visible through
the ring. This applies to horizontal or vertical bitewing holders, and periapical
(anterior or posterior) holders.
4.
Apply a new disposable sheath to the image sensor, making sure that it covers the
portion of the cable which may come into contact with the patient.
CAUTION
When twisting the disposable covers around the sensor wire, be sure not to
twist the wire.
5.
Attach the sensor to the positioner. The sensor will only fit into the holder one way.
English
Use
6-1