Shown above is the X-ray sensor DQE as a function of spatial frequency, for a range of X-ray
doses from 50 to 600 uGy. The curves indicate that there is no significant change in DQE
performance going from 50 uGy to 600 uGy. This implies that the sensor operates quantum-
limited over that dose range, indicating that all photons are being used effectively and that
noise does not degrade performance at low doses. The curve also indicates that device can
resolve detail down to 20 lp/mm (over all doses). This means that the X-ray sensor exhibits no
reduction in resolution over the input dose range from 50 to 600 uGy.

Product Symbols

BF type device (IEC 601.1 - 1988 and Amendments)
Please refer to the written instructions in this manual.
DC Current (IEC 601.1 - 1988 and Amendments)
This ETL listed mark guarantees that Intertek has certified the product
described herein under control number 3187969 to be in compliance with
the applicable regulations. Intertek is:
The CE symbol ensures that the product herein specified meets the
provisions of European Council Directive 93/42 EEC concerning
medical devices.
Manufacturer
Date of manufacture
Catalog number
Serial number
Indicates the product should be used only once.
The symbol is found on the packaging for sheaths.
English
a Nationally Recognized Testing Laboratory by the Occupational
Safety and Health Administration (OSHA) in the United States.
a Certifying Body in Canada by the Standards Council of Canada.
Specifications and Standards