TURNING ON/OFF
To turn on scale, press and hold the Zero button for
one second. The scale performs a display test and
momentarily displays the software version before
entering the active weighing mode.
To turn off the scale, press and hold the Zero button until
OFF is displayed.
CALIBRATION OPERATION
When the scale is operated for the first time, a span
calibration is recommended to ensure accurate
weighing results. Before performing the calibration, be
sure to have the appropriate calibration weights. Ensure
that the security switch is set to unlocked position.
Press and hold Menu until [MENU] is displayed. When the
button is released, the display will show [C.A.L]. Press Yes
to accept. [SPAN] will then be shown. Press Yes to begin
the span calibration.
[--C--] will be displayed while zero reading is stored.
Next, the display shows the calibration weight value.
Press the No key to toggle the value. Place the specified
calibration weight on the pan. [--C--] will be displayed
while the reading is stored. The display will show [donE]
if the calibration was successful. The scale will return to
the previous application mode and is ready for use.
WEIGHING MODE
This mode is the factory default setting.
1.
If needed, press and hold Mode until [WEIGH] is
displayed.
2. If required, place an empty container on the pan
and press Tare.
3. Add sample to the pan or container. The display will
show the weight of the sample.
COUNTING MODE
This mode counts large numbers of items based on the
weight of a reference count.
1.
If required, place an empty container on the pan
and press Tare.
2. Press and hold Mode until [Count] is displayed.
[CLr.PW] (clear average piece weight, APW) will then
display. If no APW exists, the scale will display [PUt.10],
proceed to step 5.
3. Press No to use the stored APW. Proceed to step 7.
4. Press Yes to establish an APW.
PAGE 6 OF 53
OPERATION CONTINUED
5. The scale will then display the sample size, i.e.
6. Put the indicated number of pieces on the pan and
7. Add additional pieces until the desired count is
8. To clear the stored APW, press and hold Mode
PERCENT MODE
This mode measures the weight of a sample as a
percentage of a reference weight.
1.
2. Press and hold Mode until [PErcnt] is displayed.
3. Press No to use the stored reference weight and
4. Press Yes to establish a new reference. Scale will now
5. Add the desired reference material to the pan or
6. Replace the reference material with the sample
7. To clear the stored reference, press and hold Mode
CHECK MODE
Use this mode to compare the weight to a target weight
range. The scale supports positive, negative and zero-
check weighing.
CHECK WEIGHING
Use this mode to compare the weight of items to a
target weight range.
[PUt 10]. Press No or Back to toggle the choices
(5, 10, 20, 50 or 100).
press Yes to calculate the APW. The display will show
the piece count.
reached.
until [Count] is displayed. Press Yes when [Clr.pW] is
displayed.
NOTE: Press Function to view the current APW.
If required, place an empty container on the pan
and press Tare.
[CLr.rEF] will then display. If no reference weight
exists, the scale will display [Put.rEF]; proceed to
step 5.
proceed to step 6.
display [Put.rEF].
container. Press Yes to store the reference weight.
The display will show 100%.
material. The display will show the percentage of the
sample compared to reference weight.
until [PErcnt] is displayed. Press Yes when [CLr.rEF] is
displayed.
NOTE: Press Function to view the current
reference weight.
1021 IH-9560