Sensor Inspection; Electrical Safety; Mains Isolation - KaVo DEXIS Titanium 1.014.6083 Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 177
DEXIS Titanium

Sensor Inspection

Always inspect the sensor, cable, and positioning devices for physical damage prior
to every use.
WARNING
Remove sensor from service if damage to the cable or housing is
observed, otherwise exposure to elevated surface temperature or
electrical energy may result.

Electrical Safety

The product must be used only in rooms or areas which comply with all laws and
regulations applicable to electrical safety in medical premises, where provisions for
safety ground connections are present.
All X-ray equipment for Dental Intra-oral Radiography used with the DEXIS Titanium
sensor must conform to IEC 60601-2-65.
The DEXIS Titanium sensor conforms to safety standard IEC 60601-1.
The DEXIS Titanium sensor is not suitable to be operated in oxygen rich and/or
explosive environments.
All IT components electrically connected to the DEXIS Titanium sensor must
conform to IEC 60950-1.
Normally, the IT components are placed OUTSIDE the patient environment. IT
components placed INSIDE the patient environment, due to customer site
requirements, must also conform to IEC 60601-1.
IEC 60601-1 defines the "Patient environment" as "any volume in which intentional
or unintentional contact can occur between a Patient and parts of the ME Equipment
or ME System or between a Patient and other persons touching parts of the ME
Equipment or ME System."

Mains Isolation

Disconnection from the supply mains occurs at the input to the computer. The
sensor can also be disconnected from the computer.
2-2
English

Publicidad

Tabla de contenido
loading

Tabla de contenido