DEXIS Titanium
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The DEXIS Titanium sensor, used with a compliant computer, is intended for use in the electromagnetic
environment specified below. The customer or the user of the DEXIS Titanium sensor should assure that it
is used in such an environment.
Immunity Test IEC 60601 Test
Level
Electrostatic
+ 8 kV contact
discharge
+ 2 kV, + 4 kV,
(ESD)
+ 8 kV, + 15 kV air
IEC 61000-4-2
Electrical fast
+ 2 kV for power supply lines
transient/burst
+ 1 kV for input/output lines
IEC 61000-4-4
Surge
+ 1 kV differential mode
IEC 61000-4-5
+ 2 kV common mode
Voltage dips,
0% U
T;
o
short interrup-
180
, 225
tions and volt-
0% U
; 1 cycle
T
age variations
on power supply
70% U
T
input lines
60 Hz respectively
IEC 61000-4-11
Single phase: at 0
0% U
T;
60 Hz respectively
Single phase: at 0
VDI specs per 3
<5% U
40% U
T
70% U
T
<5% U
Power frequen-
30 A/m, 50 Hz or 60 Hz
cy (50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE 1 U
is the a.c. mains voltage prior to application of the test level.
T
(*) Computer used with the DEXIS Titanium sensor must meet this rating.
B-4
o
o
o
0.5 cycle at 0
, 45
, 90
, 135
o
o
o
, 270
, and 315
; 25/30 cycles for 50 Hz and
o
250/300 cycle for 50 Hz and
o
rd
edition:
(95% dip in U
) for 0.5 cycle
T
T
(60% dip in U
) for 5 cycle
T
(30% dip in U
) for 25 cycle
T
(95% dip in U
) for 5 s
T
T
Compliance
Electromagnetic Environment -
Level
Guidance
Floors should be wood, concrete or
Complies
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Complies (*) Mains power quality should be that
of a typical commercial or hospital
environment.
Complies (*) Mains power quality should be that
of a typical commercial or hospital
environment.
o
,
Complies (*) Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the DEX-
IS Titanium sensor requires contin-
ued operation during power mains
interruptions, it is recommended
that the DEXIS Titanium sensor be
powered from an uninterruptible
power supply or a battery.
Complies
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical com-
mercial or hospital environment.
English