WARNING
Do not autoclave sensor. Autoclave sterilizers will permanently
damage the DEXIS Titanium sensor.
Recommended Disinfecting Solutions
Decontaminate the sensor, cable and holders in accordance with CDC – or your
country's (e.g., OSAP) standards – for infection control. DEXIS recommends
wiping the DEXIS Titanium sensor with one of the following cleaning agents.
CaviCide
•
CaviWipes
•
ProSpray
•
Sani-Cloth
•
DO NOT:
Clean the sensor using non appropriate tools.
•
Maintenance
DEXIS Titanium does not require any special maintenance other than regular
cleaning and disinfection. It is recommended use KaVo Software Manager for
automatic software updates.
Go to KaVo.com/downloads for more information on using KaVo Software Manager.
DTX Go may also be used for managing your software updates.
Image Quality Assurance
Image quality of the DEXIS Titanium sensor depends on several factors:
the quality of the X-ray source (kV, focal spot size, distance)
•
the alignment of the X-ray source to the anatomic region
•
the applied X-ray dose / exposure time
•
the settings of the computer monitor
•
It is recommended that you establish a procedure for periodic review of the image
quality. If image quality is not satisfactory, or degrading, please check the
contributing system parts as outlined below: