Attaching Connecting Cable And Product; Connect Up The Product And Select Unit Settings; Cleaning, Disinfection, Sterilization; Safety Instructions - Erbe 20191-395 Indicacion De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 34
All manuals and user guides at all-guides.com
OBJ_DOKU-25875-006.fm Seite 17 Dienstag, 29. November 2016 12:58 12
5.3

Attaching connecting cable and product

Use one of the connecting cables specified in these instructions
only. The connecting cable must be sterile.
Inspect the connecting cable for damage, particularly the insula-
tion. Do not use it if it is damaged.
Connect the proximal end of the instrument/the instrument
handle and the connection cable so that you notice that they
snap together.
5.4

Connect up the product and select unit settings

Plug the connection cable into the monopolar socket of the RF
surgical instrument.
Only select unit settings that do not exceed the maximum elec-
trical capacity of the product (HF peak voltage/V
voltage that the unit emits depends on the mode used and the
power selected. You can find information about the technical
data of the unit settings in the user manual of the electrosurgical
unit used.
6

Cleaning, disinfection, sterilization

6.1

Safety instructions

Not suitable for manual cleaning and disinfection.
Suitable for precleaning using an ultrasonic bath.
Never use sharp objects for cleaning.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 2 bar.
Maximum water pressure during rinsing: 2 bar.
max). The HF
p
17

Publicidad

Tabla de contenido
loading

Tabla de contenido