IMPORTANT:
2.
Wash your hands and the test site with warm water and soap to ensure
accurate results.
Thoroughly dry your hands and the test site.
Do not use lotion or cream on the test site.
3.
Check test strip expiration date.
Do not use expired test strips; they may produce inaccurate results.
IMPORTANT: Use only FreeStyle Lite® test strips. Other test strips can
produce inaccurate results. The test strips are for single use only.
11
Contact Customer Care Toll Free at 1-888-522-5226 www.MyFreeStyle.com
ART22813-003_rev-A.indd 11-12
All manuals and user guides at all-guides.com

Important Test Strip Information

Please read the FreeStyle Lite® test strip package insert for more information.
Use the test strips within the meter's operating range: 40° – 104° F (4° – 40° C).
Store test strips in a cool, dry place between 40° – 86° F (4° – 30° C).
Store test strips away from direct sunlight and heat.
Store test strips in their original vial only.
Clean and dry your hands before removing a strip from the vial.
Close the vial cap tightly immediately after removing a test strip.
Do not transfer test strips to another container.
Do not store test strips outside the vial.
Do not bend, cut, or alter test strips.
Do not apply blood to both sides of the test strip.
Do not press the test strip against the test site.
Do not scrape the blood onto the test strip.
Do not apply blood to the flat side of the test strip.
Do not apply blood to the test strip when the test strip is out of the meter.
Do not put blood or foreign objects into the meter.
Warning: Drying agents in the test strip vial may be harmful if inhaled or
swallowed and may irritate skin and eyes.