The HomeFill II Oxygen system consists of an IOH200AW gas
compressor, connection tubing, and the Invacare supplied
portable gas cylinders with the integral flow regulator and
interlocking fill fitting.
The IOH200AW compressor consists of a multistage
compressor, internal tank, control electronics with an integral
oxygen concentration sensor, interlocking fill fitting, inlet
fitting, and drive motor. Compressor operation is controlled
by the electronics assembly. The motor/compressor is allowed
to fill a cylinder only when the oxygen concentration of the
inlet gas stream is greater than 90%. The motor/compressor
fill operation is completed when the gas in a cylinder reaches
the maximum fill pressure of 2000 PSI (13786 kPa).
The IOH200AW compressor can fill cylinders for portable use
by ambulatory patients. Invacare supplied portable gas
cylinders have an interlocking fill fitting that matches only
with the Invacare IOH200AW compressor assembly. The gas
cylinder also contains an integral flow regulator with fixed,
selectable flow settings.