SECTION 6—TROUBLESHOOTING/
CYLINDER FILL TIMES
PROBLEM
There are no indicator
lights on when
compressor is turned on.
RED light On
Compressor cycles On
and Off
or
O
Below Normal -
2
alternating (YELLOW)
and Filling (GREEN) lights
or
O
Below Normal
2
(YELLOW) light goes
On and Off
Part No 1145805
TROUBLESHOOTING
1. Turn compressor Off (O).
2. Check that the power cord is correctly plugged
into compressor and wall socket.
3. Turn the compressor on ( ).
4. Clean air filters. Refer to Maintenance on
page 36.
5. If performing STEPS 1-4 does not cause any
indicator lights to come on, contact your home
health care provider or Invacare for service.
1.Check the cylinder gauge to see if it is full.
Cylinder full - Remove the cylinder and turn the
unit Off (O). Perform the Cylinder Prefill
Inspection on page 19 with a cylinder that is not
full.
Cylinder not full - Proceed to STEP 2.
2. Check that the cylinder is properly connected.
Turn the compressor Off (O). Remove the
cylinder and reinstall. Turn the compressor On
( ).
3. If the cylinder is not full and is properly
connected, the compressor may have an internal
failure.
Contact your home health care provider or
Invacare.
1. Check the concentrator liter flow setting is at
2.5 L/min. or less. Where necessary, adjust the
litre flow setting (refer to the concentrator
owner's manual). Wait at least 20 minutes for the
GREEN light to appear. If it does not appear,
proceed to STEP 2.
2. The concentrator may need service. Contact
your home health care provider or Invacare.
33
SOLUTION
HomeFill
II Compressor
®