Error which has occurred
Possible causes
Cuff was applied incorrectly or too loosely.
Measurement was disturbed due to speaking
or moving.
The pulse could not be measured accurately.
The battery is empty. Replace the battery.
The device has a technical problem. Contact
the shop where you bought the device or the
HARTMANN representation for your country.
Cuff will not inflate
Cuff connection is incorrectly positioned in
device.
Implausible reading
Incorrect cuff size.
Cuff placed on top of clothing.
Rolled up clothing impedes blood circulation.
Moving, talking or excitement during meas-
urement.
Deep breathing during measurement
process.
No relaxation before measurement.
Stimulants taken just before measurement.
Switch the device off if an error symbol appears. Check the possible causes
and note the instructions on self-measurement in Chapter 2. Relax for a
minute and take the measurement again. Do not move or talk during the
measurement.
14
All manuals and user guides at all-guides.com
12. Technical specifications
Measuring method:
Display range:
Measuring range:
Technical accuracy of
measurement:
Clinical accuracy of
measurement:
Power supply:
Automatic cut-out:
Cuff:
Storage capacity:
Operating conditions:
Storage/transport conditions: Ambient temperature: – 20 °C to + 60 °C
Serial number:
English
oscillometric
0 – 280 mmHg
Systole (SYS): 50 – 270 mmHg
Diastole (DIA): 30 – 250 mmHg
Pulse:
40 – 200 Pulses/min.
cuff pressure: +/– 3 mmHg
Pulse: +/– 5 % of displayed pulse frequency
complies with the EN 1060 standard, part 3.
4 x 1.5 V Mignon R6/LR6
3 minutes after end of measuring process.
Normal cuff 22 – 32 cm
1 measurement
Ambient temperature: +10 °C to + 40 °C
Relative humidity:
30 – 85 %
Relative humidity:
10 – 95 %
see battery compartment