6.2
Cleaning
Clean the device using an alcohol-based cleaning agent.
To clean the device:
1. Apply the cleaning agent to a disposable cloth and wipe the
device. Do not apply cleaning agent directly to the device.
Make sure that no liquid enters the gas inlet.
2. After cleaning, store the device without its packaging at room
temperature (20–25°C) for at least 24 hours.
3. Then check that the there is no residual alcohol inside the device
via passive sampling and checking that the result NO ALC. is
displayed.
7
Storage
● Do not permanently store the device in its delivery packaging.
NOTICE
Risk of damaging the device!
Leaking battery acid can damage the device.
► Remove the batteries if the device is not used for a prolonged
period of time.
Maintain the charge level of the rechargeable batteries.
For nickel-metal hydride batteries:
To ensure the operational readiness of the device, fully charge the
rechargeable battery at least every three months, even if it is not
used.
For lithium-ion batteries:
Charge to at least 50% every 6 months if not used for a prolonged
period of time to avoid shortening the life span.