Press the "START Ph" pushbutton to start the test. The orange
"START Ph" LED flashes for approximately 10 seconds. Step
1 must be performed during these 10 seconds.
If the "
" LED lights up, stop the test (see 3.2).
Step 1:
Place the red touch prod on the first phase and the black touch
prod on the second phase (after pressing the triggers) and
wait for the orange "START Ph" LED to remain lit constantly
without flashing.
Step 2:
Move the touch prod onto the third phase.
Result:
if the phase rotation order is direct , the red LED lights up and
the orange led goes off.
if the phase rotation order is reversed, the red LED remains
off and the orange LED goes off. In this case, retest the phase
rotation order after switching round the two leads in step 1
and move on to step 2. The red LED should light up.
NB: The voltage detection function remains active during this test.
Specifications and field of use
Voltage detection: 50 V < V < 400 V
Frequency: 50 Hz and 60 Hz (+/- 5%)
Response time: < 500 ms after each contact.
8. GENERAL SPECIFICATIONS
8.1
Power supply
3 x 1.5 batteries (R6 or LR6 type).
Battery charge life: 10,000 measurements each lasting 10
seconds (with LR6 batteries).
Battery charge indication: see 3 - Autotest
8.2
Buzzer
Repeated buzzes (2 kHz) for testing for the presence of a
voltage between 50 V and 230 V.
Continuous buzzing (2 kHz) for test of voltage > 400 V.
Typical sound level: > 60 dB(A)
8.3
Duration of operation at 400 V
For 10 consecutive ON/OFF cycles
ON < 30 seconds
OFF > 240 seconds
8.4
Climatic conditions
Reference conditions: 0°C to +30°C
Use:
temperature: -10°C to +55°C.
relative humidity: < 96% RH.
altitude: up to 2,000 m.
Storage: temperature: -40°C to +70°C (without battery)
17