With
the
RS
232
communication is established by selecting HOST. Data
are sent to an external computer every time the
external computer asks for the values.
The data available are:
•
Speed adjust
•
Pump/controller operating condition
•
Cycletime
•
Pumplife
•
Pump temperature
•
Pump current
•
Pumpvoltage
•
Controller output frequency
•
Cyclenumber
•
R1 condition
•
R2 condition
•
Life time and cycle # zeroing
•
Configuration parameter readings
•
Configuration parameter setting
Note that the new input value are put in effect only at
the next STOP/START of the pump.
If PRINT is selected and a printer is connected on RS
232
line,
an
unidirectional
established and every minute the data are sent to the
printer, even if the pump is not running.
The set of data available are:
•
Pumpspeed KRPM
•
Pump temperature
•
Pumpcurrent A
•
Pump power W
•
R1 condition
•
R2 condition
87-900-873-01 (A) - TURBO-V 550 CONTROLLER
connected,
a
bi-directional
communication
The controller is factory-set to HOST.
•
Confirm the selection by pressing the PUMP
CURRENT push-button.
•
In this way you enter into an operating phase
named "Monitor Relay Programming" described in
the following paragraph.
Monitor Relay Programming
•
The display shows:
S P E E D
S E L E C T
where: XXKRPM = is the switch point of relay
the preset turbopump speed, adjustable from 00 to 99
KRPM.
The speed threshold will condition the R1 and
operation (see the following cycle diagram] and it is
factory-set to 40 KRPM.
Press the CYCLE NUMBER push-button to select the
first number.
is
S P E E D
S E L E C T
TECHNICAL INFORMATION
T H R E S H O L D
I O N : X X K R P M
T H R E S H O L D
I O N : X X K R P M
R1 at
R2