Electrode Cable Compatibility
All of Snowden-Pencer monopolar electrosurgical devices are compatible with the
Unipolar Bovie Cord with universal plug (88-9199) or equivalent.
Pre-processing Instructions
Initiate cleaning of device within 2 hours of use.
Transport devices via the institutions established transport procedure.
Remove excess gross soil as soon as possible after use by rinsing or wiping the
device.
All devices must be processed in the completely open configuration.
All flush ports shall remain in the fully open position.
Manual Cleaning
(Steps 5, 6, 7, 9 and 11 are required for lumen devices only.)
1. Ensure all pre-processing instructions are followed prior to cleaning.
2. Prepare the enzymatic/neutral pH detergent solution, utilizing drinking/
potable water with a temperature range of 27°C to 44°C (81°F to 111°F), per
manufacturer's instructions.
3. Place the device in the open/relaxed position, with flush port open. Completely
immerse the device in the detergent solution and allow it to soak for a minimum
of 5 minutes. Actuate all movable parts during the initiation of the soak time.
4. Using a soft bristled brush, remove all visible soil from the device. Actuate the
device while brushing, paying particular attention to hinges, crevices and other
difficult to clean areas. Note: It is recommended that the detergent solution be
changed when it becomes grossly contaminated (bloody and/or turbid).
5. For lumen devices, use a soft bristled brush with a brush diameter and length
that is equivalent to lumen diameter and length. Scrub the lumen (i.e. angulated/
nonangulated positions) until no visible soil is detected in the lumen rinsing step
below.
6. For lumen devices, place the device into the open/relaxed position with the
distal tip pointed down. Flush the device with a minimum of 50 ml of detergent
solution utilizing a temperature range of 27°C to 44°C (81°F to 111°F), by
using the flushing port located on the handle/shaft. Repeat the flush process a
minimum of 2 times (i.e. total of 3 times); ensuring all fluid exiting the lumen is
clear of soil.
7. For lumen devices, if visible soil is detected during the final lumen flush,
re-perform brushing and flushing of the lumen.
Note: The final rinse steps (8 and 9) should be performed with treated water:
deionized, distilled or reverse osmosis.
8. Rinse the device by completely immersing it in treated water with a temperature
range of 27°C to 44°C (81°F to 111°F), for a minimum of 30 seconds to remove
any residual detergent or debris.
9. For lumen devices, following the rinsing step, place the device into the open/
relaxed position with the distal tip pointed down. Flush the device with a
minimum of 50 ml of treated water utilizing a temperature range of 27°C to 44°C
(81°F to 111°F), by using the flushing port located on the handle/shaft. Repeat
the flush process a minimum of 2 times (i.e. total of 3 times).
10. Dry the device with a clean, lint-free towel.
11. For lumen devices, manipulate the device to allow rinse water to drain from the
lumen.
12. Visually examine each device for cleanliness.
13. If visible soil remains, repeat cleaning procedure until the device is thoroughly
clean.
3