Automatic Cleaning
1. Ensure all pre-processing instructions are followed prior to cleaning.
2. Clean the devices via the Automatic cleaning parameters below.
PHASE
RECIRCULATION TIME
Pre-Wash
Enzyme Wash
Wash
Rinse
Pure Rinse
Drying
3. For lumen devices, manipulate the device to allow rinse water to drain from the
lumen.
4. If visible moisture is present, dry the instrument with a clean, lint-free towel.
5. Visually examine each instrument for cleanliness.
6. If visible soil remains, repeat cleaning procedure until the device is thoroughly
clean.
Inspection/Maintenance
Proper care and handling is essential for satisfactory performance of any surgical
device. The previous cautions should be taken to ensure long and trouble-free
service from all your surgical devices. Inspect devices before each use for broken,
cracked, tarnished surfaces, movement of hinges, and chipped or worn parts. If any
of these conditions appear, do not use the device. Return devices to an authorized
repair service center for repair or replacement.
Before sterilizing, lubricate the device with instrument milk or a steam permeable/
water soluble lubricant, following the lubricant manufacturer's instructions.
Immediately proceed to recommended steam sterilization. Do not store instrument
prior to sterilization.
Let devices drip dry for three (3) minutes before packaging for sterilization.
Replace thumb rings (part number SP90-1111) when worn or discolored.
Packaging
Devices can be loaded into dedicated packaging systems. Sterilization wrap material
must be cleared for the applicable sterilization modality by your country's regulatory
body. Use in accordance with packaging manufacturer's sterilization instructions
being sure to protect jaws and cutting edges from damage.
MINIMUM
15 Seconds
1 Minute
2 Minutes
15 Seconds
10 Seconds
N/A
WATER
TEMPERATURE
Cold Drinking/Potable
Water
1°C - 16°C
(33°F - 60°F)
Hot Drinking/Potable
Water
43°C - 82°C
(110°F - 179°F)
Hot Drinking/Potable
Water
43°C - 82°C
(110°F - 179°F)
Hot Drinking/Potable
Water
43°C - 82°C
(110°F - 179°F)
Purified Water (Hot)
43°C - 82°C
(110°F - 179°F)
N/A
4
DETERGENT TYPE AND
CONCENTRATION
(IF APPLICABLE)
N/A
• Detergent: pH-neutral/
enzymatic
• Concentration: Per the
detergent manufacturer's
recommendations
• Detergent: pH-neutral
cleanser
• Concentration: Per the
detergent manufacturer's
recommendations
N/A
N/A
N/A