Do not inflate to above 300 mmHg !
Measurement time should be ca. 2 minutes maximum.
Between 2 measurements allow an interval of at least 2
minutes.
The cuff must be selected such that it fits the arm circum-
Please use only boso cuffs.
ference.
The patient must follow these basic rules when having
his/her blood pressure measured:
– sit comfortably
– not cross his/her legs
– rest his/her back and arms on the chair or another surface
– have the centre of the cuff placed on the upper arm at the
level of the heart
– relax as much as possible and keep silent while his/her
blood pressure is being measured
– be calm for 5 minutes before the first reading
The following phases are recommended in the case of aus-
cultatory measurement
– adults: phase V (K5)
– children aged 3 to 12: phase IV (K4)
– pregnant women: phase V (K5), except when sounds are
audible when the cuff is deflated: in this case phase IV
(K4) Korotkoff sounds should be used.
K5 is the point at which tones heard through the stethoscope
are no longer audible
K4 is the point at which tones heard through the stethoscope
change from a clear sound to a muffled sound.
Inscriptions:
"mmHg" is the unit in which blood pressure is measured
"shock protected" = the instrument has been tested and fo-
und to work correctly after being dropped from a height of up
to one metre.
Fitting the Cuff:
The cuff must not be fitted over wounds as this could lead to
further injury.
Blood pressure must always be measured on the arm with
the higher pressure values.
7