DATA TRANSMISSION

With the advanced Bluetooth 4.0 technology applied, the mobile or portable
equipments, which are equipped with Bluetooth function in line with BLE Technical
Specifications as well as BLP Protocol established by global organization Bluetooth
SIG, are capable to receive your personal health data.
When both the Blood Pressure App and Bluetooth are ON, the monitor will
transmit measurement data to your mobile device via Bluetooth.
If the stored measurements are not transmitted, when you recall the memory,
the Bluetooth icon blinks. Please turn on the app and Bluetooth, the data will be
transmitted. The Bluetooth icon will disappear when the data is transmitted.
Bluetooth Module No.: AW2540MV1
RF Frequency Range: 2402 MHz to 2480 MHz
Output Power Range: -1 dBm
Supply Voltage: 2V-3.6 V
Transmitting Distance: 10 meters
CAUTION
Interference may occur in the vicinity of equipment marked with the following symbol:
Sensitive people, including pregnant women pre-eclamptic and those who implanted medical
electronic instruments, should avoid using the unit whenever possible.
Keep the monitor at least 20 centimeters away from the human body (especially the head)
when the data transmission is proceeding after measurement.
To enable the data transmission function, this product should be paired to Bluetooth end at 2.4
GHz.
How to mitigate possible interference?
1.
The range between the device and BT end should be reasonably close, from 1 meter to
10 meters. Please ensure no obstacles between the device and BT end so as to obtain
quality connection and to lower the RF output range.
2.
To avoid interference, other electronic devices (particularly those with wireless
transmission / Transmitter) should be kept at least 1 meter away from the monitor.
ENGLISH • 19