CONTROLLER SPECIFICATIONS
Input:
Voltage
Frequency
Power
Output:
Voltage
Frequency
Power
Operating
temperature
Storage temperature
19, 18, 69 Vac ±10%,
1-phase via an external
transformer
47 to 63 Hz
350 VA maximum
54 Vac nominal ±10%,
3-phase
1250 Hz, ±2%
150 W maximum
0 °C to +40 °C
-20 °C to +70 °C
P1 optoisolator input
J2 optoisolator output
P6 output fan
Weight:
CONTROLLER OUTLINE
The outline dimensions for the Turbo-V 70 PCB
controller are shown in the following figure.
Controller outline
50
TECHNICAL INFORMATION
Minimum ON
3mA
Maximum
5mA
24 Vdc
60mA
100 mA
0.5 kg (1.1 lbs)