Descargar Imprimir esta página

H&O Equipments CryoPen b+ Manual De Usuario página 12

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Step 2
I
Load a new cartridge
The cartridges are packed in peel-pack blisters with an assembled filter
and cartridge opener included. The filters are used for protection
of the applicator. The opener punctures the cartridge.
CryoPen B+ can be used with 23,5g cartridges only. Ensure that the
loading cavity is empty of all residue.
Open the peel-pack. Take out the filter/opener with the tweezers.
Hold the CryoPen's tip end pointed downward. Place the filter in the
instrument and make sure that it is seated flat at the bottom
of the loading cavity.
Step 3
I
Puncture the unused cartridge
Gently slide the cartridge into the CryoPen's loading cavity and
start screwing the cartridge turning it clockwise in a quick and
uninterrupted movement until hand-tight.
12
Installation of the assembled filter/opener device