EN-8
4. APPLICATIONS
The PX balance can be operated in 6 application modes by long pressing the Function / Mode button.
4.1
Weighing
Note: Before using any application, be sure the balance has been leveled and calibrated.
Use this application to determine the weight of objects in the selected unit of measure.
Weighing
1. Press Tare or Zero if necessary to begin.
2. Press and hold the Function / Mode button to select
Weighing (this application is the default).
3. Place objects on the pan to display the weight. Once
the reading is stable, the * will appear.
4. The resulting value is displayed in the active unit of
measure.
Item Settings
To view or adjust the current settings.
•
Capacity Bar: When set to On, the capacity bar is displayed in the reference field. The capacity will not
display when the balance is set to zero.
•
Weighing Units: Change the displayed unit. See Section 5.4 for more information.
•
Filter Level: Change Filtering level. See Section 5.3.2 for more information.
•
GLP Data: See Section 5.7 for more information.
•
Print Settings: Change printing settings. See Section 7 for more information.
4.2
Parts Counting
Note: Before using any application, be sure the balance has been leveled and calibrated. The minimum piece
weight should be no less than 0.1d.
Use this application to count samples of uniform weight.
Parts Counting
1. Press Tare or Zero if necessary to begin.
2. Press and hold the Function / Mode button until
Parts Counting appears.
PX Series Balance