EN-20
Item Settings:
Sample Type: Liquid
Volume: The calibrated sinker has a default volume of 10.0 ml,
which can be increased or decreased by pressing No or Back.
After setting the volume, press Yes to begin the weighing.
Note: when the Density Type is set to Liquid, the Liquid type
and Porous material selections are disabled.
1. Weigh the calibrated sinker in the air with the balance and
the density determination kit.
When the * (symbol of stability) appears, press the
Function / Mode button to confirm the weight of the
calibrated sinker in the air.
2. Weigh the calibrated sinker submerged in the liquid with
the balance and the density determination kit. Lower the
calibrated sinker down into the liquid until it is fully
submerged (1 cm below the surface of the liquid).
3. When the * (symbol of stability) appears, press the
Function / Mode button to confirm the weight of the
calibrated sinker. The density of the liquid is displayed.
4. After completion of the test, press the Function / Mode
button to test a new sample.
PX Series Balance