4 Installation and connection information

4 Installation and connection information
4.1

General information

4.2

Terminals and jumpers

Minebea Intec
-
Only use the cable glands provided by the manufacturer.
-
Install the junction box so that the cable glands are on the bottom.
-
Do not open the junction box when connected to the voltage supply.
Note:
At ambient temperatures >45°C, cables that are suitable for at least 85°C must be used.
Signal +
green
Signal -
gray
Supply +
red
Weighbridge Cable Junction Box PR 6021/08
Terminal contacts
+ measuring voltage (load cells)
1–8
Terminal contact
+ measuring voltage (connection cable)
IND
Terminal contacts
- measuring voltage (load cells)
1–8
Terminal contact
- measuring voltage (connection cable)
IND
Terminal contacts
+ supply voltage (load cells)
1–8
Terminal contact
+ supply voltage (connection cable)
IND
Terminal contact S + sense (connection cable – white)
EN-11