Procedure - Minebea Intec PR 6021/08 Manual De Instalación

Ocultar thumbs Ver también para PR 6021/08:
Tabla de contenido

Publicidad

Weighbridge Cable Junction Box PR 6021/08
5.5.2

Procedure

EN-18
NOTICE
Loss of calibration accuracy
An unstable signal can arise due to contact problems. This results in luctuating zero
points, which lead to the loss of calibration accuracy.
Do not use a potentiometer.
Only use resistors of 0–5.62 Ω (1%, P70 = 0.6 W) of size MBB0207
(approx. 2.5 × 6.5 mm) or CECC B.
Electronic corner correction allows the sensitivity of each individual load cell to be
reduced separately.
The resistor required here can be calculated using the following formula:
R
Weight
actual
Weight
set point
R
LC
Example:
placed load
Display
R
LC
Calculated resistance
The following steps are necessary to minimize corner load errors:
1.
Select the load cell with the lowest displayed load as the reference cell.
2. In the junction box, remove the jumper from the relevant supply line (Supply +) and
solder in a resistor according to the abovementioned formula (in the
example 4.68 Ω).
This aligns the load cells with the displayed value of the reference cell.
Note:
The soldering studs are factory short-circuited for the resistances.
3. After that, recalibrate the scale.
Resistor integrated in the input circuit of the load
cell (the resistance value is normally <7 Ω).
Weight value on the display
Weight of the placed load
Input resistance of the load cell
12,000 kg
12,052 kg
1080 Ω (see Installation manual of the load cell)
4.68 Ω
5 Cable connections
Minebea Intec

Publicidad

Tabla de contenido
loading

Tabla de contenido