Troubleshooting - Freestyle Freedom Lite Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Troubleshooting

What It Means
1. The meter
Test strip is inserted
does not
upside down, or bottom
enter test
end in, or not fully
mode after
inserted into the meter
inserting a
test strip.
No battery is installed
Battery is installed
incorrectly
Dead battery
Defective test strip or
meter
Blood or foreign objects
put in the test strip port
21
Contact Customer Care Toll Free at 1-888-522-5226 www.MyFreeStyle.com
ART22812-003_rev-A.indd 21-22
All manuals and user guides at all-guides.com
What To Do
1. Insert test strip print side up, top end in.
1. Install battery with (+) facing up.
1. Replace battery. Reset date and time, if
necessary.
2. If meter still does not enter test mode,
contact Customer Care.
1. Contact Customer Care.
What It Means
Blood drop is too small
1. Add blood to the same sample area within
2. The test
60 seconds, or repeat the test using a new
does not
test strip and a larger blood drop.
start after
applying
the blood
Defective test strip
1. Repeat the test using a new test strip. Wait for
drop.
Sample applied after
meter turns off
or control solution.
2. If problem persists, contact Customer Care.
Defective meter
1. Contact Customer Care.
What To Do
and
to appear before applying blood
22
11/10/16 11:02 AM