Limitations:
•
For in-vitro diagnostic use only (external use only).
•
Not for diagnosis of or screening for diabetes.
•
Not for testing of neonates.
•
The test strips are for single use only. Do not reuse test strips.
•
The system can give false results greater than 10,000 feet (3,048 meters) above sea level.
•
The system can give false results outside the hematocrit range: 15% to 65%.
•
Individuals experiencing a hypoxia state may yield falsely low results.
•
In situations of decreased peripheral blood flow, capillary blood testing may not be
appropriate as it may not represent the true physiological state. Examples would include but
are not limited to: severe dehydration caused by diabetic ketoacidosis or the hyperosmolar
non-ketotic state, hypotension, shock, or peripheral vascular disease.
•
This device should not be used to test critically ill patients.
•
Alternative site testing (AST) should not be used to calibrate continuous glucose monitors
(CGMs) nor for use in insulin dose calculations.
•
AST testing should only be done during steady-state times (when glucose is not changing rapidly).
•
For single patient use at home by a lay user.
•
Severe dehydration or excessive water loss may cause false low results. If you believe you are
suffering from severe dehydration, consult your physician immediately.
•
Cholesterol up to 500 mg/dL or triglycerides up to 3,000 mg/dL do not significantly affect test
results. However, glucose values in specimens beyond these levels, should be interpreted with
caution.
4
Atkins SH, Dasmahapatra A, Jaker MA, Chorost MI, Reddy S: Fingerstick glucose determination in shock. Ann Int Med 114: 1020-
1024. 1991.
5
Sandler M, Low-Beer T: Misleading capillary glucose measurements. Practical Diabetes 7: 210. 1990.
Wickham NWR, Achar KN, Cove DH: Unreliability of capillary blood glucose in peripheral vascular disease. Practical Diabetes 3: 100. 1986.
6
3
Contact Customer Care Toll Free at 1-888-522-5226 www.MyFreeStyle.com
ART22812-003_rev-A.indd 3-4
All manuals and user guides at all-guides.com
Setting Up the Meter
Purpose
The purpose of this section is to set the meter sound volume, time and date.
4, 5, 6
1.
m
With the meter off, press and hold
for 3 seconds until
2.
Press
m
to bypass setting the optional reminder alarms. You may choose to
set these later (see Setting Reminder Alarms section).
appears.
4
11/10/16 11:02 AM