Etac R82 Wombat Living 1 Guia Del Usuario página 39

Tabla de contenido

Publicidad

Electromagnetic compatibility (EMC) environ-
ment
This equipment is for use in home healthcare facility
environment.
This equipment is not likely to cause any harmful
interference in nearby equipment; however, it can
be impaired by electromagnetic (EM) interference
emitted by other equipment, which can lead to
malfunctions. Do not use this equipment where
there is a risk that other electrical and/or electronic
equipment can interfere with this equipment.
This equipment complies with the following
standards
IEC 60601-1-2:2014 Medical electrical equipment
- Part 1-2: General requirements for basic safety
and essential performance - Collateral Standard:
Electromagnetic disturbances - Requirements and
tests (EN 60601-1-2:2015)
EN 12182:2012, Clause 7, Assistive products for
persons with disability – General requirements and
test methods
Testing for EM disturbances (EM emission and EM immunity)
Test modes during emission tests:
1. Movement mode (battery). Actuator running in/out
2. Charging mode, actuator not running.
EM emission standards
Conducted and radiated RF emission CISPR 11
Harmonic distortion
IEC 61000-3-2
Voltage fluctuations and flicker
IEC 61000-3-3
support.R82.org
EMC - WARNINGS
Use of this equipment adjacent to or stacked with
other equipment should be avoided because it
could result in improper operation. If such use is
necessary, this equipment and the other equipment
should be observed to verify that they are operating
normally.
Use of accessories, transducers and cables other
than those specified or provided by R82 could
result in increased EM emission or decreased EM
immunity of this equipment and result in improper
operation. Several power plugs are supplied with
this equipment. Please use the correct power plug
suitable for your country. Please chose among
the supplied power plug adaptor units: English,
Australian, EU, US/Japanese plugs.
Portable radio frequency (RF) communication
equipment (including peripherals such as antenna
cables and external antennas) should be used no
closer than 30 cm (12 inches) to any part of the
Wombat Living, including cables specified by R82.
Otherwise, degradation of the performance of this
equipment could result.
Classification/levels
Group 1 (the product uses RF energy only for its
internal function)
Class B (the product is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes)
Class A.
No testing performed, as the rated power for this
equipment is less than 75 W. No limits are specified
for equipment with a rated power of 75 W or less.
dmax ≤ 4 % (without additional conditions)
39
EN

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

R82 wombat living 2R82 wombat living 3

Tabla de contenido