4. Device setup

Note: For a better understanding of
each step, please see also pages 2
and 3.
4.1 Connecting the device,
performance check
1. Connect the power cord to socket (6)
of the device and the mains plug to a
wall outlet with a non-fused earthed
wire (100 to 240 Volt, 50/60 Hz).
2. Turn the power switch (7) on.
3. Follow these steps to set the carriage
to the home position:
Initial adjustment for new patients
Write the patient's name down on the
back of the chip card. Insert the original
patient chip card (2) into the program-
ming unit (1).
Press the MENU key on the program-
ming unit three times to access pro-
gramming level 3 (with each key press,
you advance one level).
Press the "new patient" parameter key
and select this function (a
check mark appears in the circle next to
the function).
Press the START key. The CPM device
automatically enters the home
position.
Setup with programmed chip card
Insert the original patient chip card (2)
into the programming unit (1).
Press the START key.
The device automatically enters the
starting position (maximum adduction,
halfway between internal / external
rotation).
Performance check
If the programming unit can be oper-
ated as described above and
OptiFlex
®
S enters the home position
(for home position values, refer to
section 5.3), the device has passed the
performance check.
The device also runs performance
checks regularly during operation. This
is what happens, if a problem is identi-
fied:
– An audio signal sounds.
– The device switches off immediately.
– The message "ERROR" and an error
code (e.g. ERROR 5) appear on the
display.
In this situation, you may attempt to
restart the device by turning it briefly
off and on again with the power
switch. If the error message persists,
have the device inspected by a
Service technician, before using it
again.
If it has been determined that the
OptiFlex
®
S device operates perfectly,
ask the patient to sit down on the
OptiFlex
®
S chair.