5. Setting the treatment values
Warning!
Patient hazard ⎯
Before therapy, a test run consisting
of several exercise cycles must be
completed without the patient. Then
repeat the test run with the patient and
check that the movement does not
cause any pain.
Note: See also 2.2 and 2.3 !
Note!
Programming is only possible when
the patient's chip card has been
inserted.
For information about therapy para-
meters and programming of special
functions, please refer to sections 5.1
See section 5.6 for programming
examples.
Important!
The programming unit OptiFlex
"Graphics" can be connected to all
products of the OptiFlex
series with serial numbers 3000 and
higher.
It is possible to exchange chipcards
between the "Text" and "Graphics"
versions.
Please note:
1. If you use a "Graphics" program-
ming unit with a formatted chipcard
of the "Text" version, please note
the following
- the selected language of the"Text"
version remains activated"; it is
irrelevant for the "Graphics"
version
®
S
®
S shoulder
18
- the reverse on load setting saved
in the "Text" version is automati-
cally used for both motors
- the key lock is disabled.
2. If you use a "Text" programming
unit with a formatted chipcard of
the "Graphics" version, please note
the following
- the selected language automati-
cally changes to the default
setting: German
- the reverse on load setting saved
for motor A in the "Graphics"
version is automatically used for
both motors
- a key lock set in the "Graphics"
version will remain active,
because it is irrelevant for the
"Text" version.
5.1 General information on
programming OptiFlex
1. You activate the programming mode
by briefly pressing the MENU key on
the programming unit.
2. The treatment parameters and
functions are allocated to four
programming levels (four per level).
To be able to program a parameter
you will have to access the corre-
sponding programming level. This is
also done with the MENU key. With
each key press you advance one
level. The code M1, M2, etc. that
appears in the middle of the display
indicates the programming level.
3. You activate the treatment parame-
ters and functions with the four
parameter keys below the display.
The symbols above the four parame-
ter keys indicate the assigned param-
eters and functions.
This is what happens when you press
®
S