LEVEL 4:
■ Transport setting
With this function, the carriage will
move to a position optimally suited
for packing the CPM device. Select
the function and press the START
key. The carriage moves to the
transport position. In the top left-
hand corner of the display you will
see
(also refer to section 6.3
Transport).
■ Synchronized/
non-synchronized mode
Motors A and B can run in synchro-
nized or non-synchronized mode.
Synchronized:
Motors A and B perform a synchro-
nized movement according to the
shoulder joint's physiological motion
pattern:
Starting from the initial position
(maximum adduction / halfway
between internal and external rotation
position), the carriage will first move
to the maximum internal rotation
position. The carriage will then
simultaneously move to the maxi-
mum abduction position and the
maximum external rotation position
and subsequently to the maximum
adduction position and the maximum
internal rotation position. After
reaching this position, the cycle
restarts: maximum abduction and
maximum external rotation.
During operation the synchronized
mode is indicated in the top left-hand
corner by the symbol
Non-synchronized:
Both motors run independently of
each other within the selected ranges
of motion.
To enable the non-synchronized
mode, disable the synchronized
mode.
During operation the non-synchro-
nized mode is indicated in the top
left-hand corner by the symbol
Default: synchronized mode
enabled
Patient hazard ⎯
The synchronized operation is recom-
mended as a general rule. However,
medical and/or therapeutic indications
may call for the non-synchronized
mode. In the non-synchronized mode,
particular caution and attention on the
part of the physician/therapist is
required to rule out any risk for the
patient.
■ Total therapy time
Under menu item "total therapy time "
you can view each patient's total
therapy time (duration of all the
patient's therapy sessions).
To delete the stored therapy time:
Press and hold the parameter key
for 5 seconds or select the "New
patient" function.
■ Service menu
For service purposes only, refer to
Service Manual.
Reminder:
You save the selected parameter values
by pressing the STOP key.
.
24
Caution!