15.10 Sterilization
Sterilization procedure:
Steam sterilizer:
Prevacuum phases:
Minimum temperature:
Minimum holding time:
Minimum drying time:
Maximum temperature:
Maximum holding time:
15.11 Storage
After sterilization, store the product dry and dust-free in the sterilization
packaging.
15.12 Additional Information
Does not apply.
16
Application Instructions
WARNING
•
The product is not sterile. Process the product before first and any further
application.
This is the only way to ensure the product is germ-free and functional.
•
Only use intact products from intact sterile packaging or sterilisation
containers. Only use products that have not expired.
This is the only way to ensure the product is germ-free and functional.
•
Do not leave any parts in the patient.
Otherwise there are risks for the health of your patient.
•
Always correctly assemble the product and all products used in combination
with it, including their components.
Otherwise there are risks for the health of your patient.
•
Do not bend the instruments by applying excessive pressure. Do not bend
back deformed instruments.
Otherwise, injury may be caused to the patient and user.
Before using the product, perform a function test.
22
Fractionated prevacuum method
Selectomat HP 666-1 HRED (MMM)
3
132 °C
4 min
20 min
137 °C
18 min