15.10 Sterilization
Sterilization procedure:
Steam sterilizer:
Prevacuum phases:
Minimum temperature:
Minimum holding time:
Minimum drying time:
Maximum temperature:
Maximum holding time:
15.11 Storage
After sterilization, store the product dry and dust-free in the sterilization
packaging.

15.12 Additional Information

Does not apply.
16

Application Instructions

WARNING
The product is not sterile. Process the product before first and any further
application.
This is the only way to ensure the product is germ-free and functional.
Only use intact products from intact sterile packaging or sterilisation
containers. Only use products that have not expired.
This is the only way to ensure the product is germ-free and functional.
Do not leave any parts in the patient.
Otherwise there are risks for the health of your patient.
Always correctly assemble the product and all products used in combination
with it, including their components.
Otherwise there are risks for the health of your patient.
Do not bend the instruments by applying excessive pressure. Do not bend
back deformed instruments.
Otherwise, injury may be caused to the patient and user.
Before using the product, perform a function test.
22
Fractionated prevacuum method
Selectomat HP 666-1 HRED (MMM)
3
132 °C
4 min
20 min
137 °C
18 min