PIP MODE
What to do
6.
Discharge liquid
7.
Start blow-out?
START BLOW-OUT MANUALLY
The blow-out function can, if necessary, be initiated manually
at any time.
1.
Bring up the
blow-out function
2.
Start blow-out
NOTE:
To accomplish the blow-out, the piston moves to its lowest position.
The user must be certain that any residual liquid is discharged safely.
If the pipetting key is pressed and held, the piston will be maintained
at its lowest position to avert an accidental aspiration of liquid.
When the key is released, the piston returns to the start position
How to do it
Press the pipetting key once
to discharge the liquid. The
arrow in the display points
downwards (discharge).
No action required!
When pipetting in the PIP
mode the blow-out function
is performed automatically.
Press the ENTER key. The
display shows 'blo'
for blow-out.
Press the pipetting key once
to initiate the blow-out pro-
cess. The display moves back
to the start position of the
selected pipetting mode.
Keys to press
Display readout
1x
1x
1x
13
ErgoOne E · User Manual ·