EASY CALIBRATION Adjustment of the ErgoOne
The calibration mode 'CAL'
Adjustment
The instrument should be set to either the nominal volume (for example
200 µl for a 200 µl pipette) or a specific test volume, in the standard
pipetting mode (PIP). See page 8 and 12 for procedures.
E.g., volume according to testing of volume 201.3 µl.
What to do
1.
Bring up the CAL
mode
2.
Activate the CAL
mode
3.
Confirm CAL mode
4.
Set the volume
5.
Confirm volume
24
· ErgoOne E · User Manual
How to do it
Press and hold the MENU
key (> 3 sec) until CAL ap-
pears. The display reads 'off'.
'CAL' blinks.
Press either the UP or DOWN
key to activate the CAL
mode. The display changes
from ‚off' to ‚on'. 'CAL'
continues to blink.
Press the ENTER key. The
display now shows the set
pipetting volume.
'CAL' blinks.
Use the arrow keys (+/-) to
set the volume, which was
previously determined and
tested. 'CAL' blinks.
Press the ENTER key.
The display shows the tested
and corrected volume. The
CAL symbol is continously
displayed to confirm that an
adjustment has been made.
E
®
Keys to press
1x
>3s
1x
1x
Display readout