InPro 86X0 i e Series
Inspection of the turbidity sensors InPro 86X0 i e
Thank you for buying the verification kit for the InPro ® 86X0 i e series turbidity sensors from
METTLER TOLEDO. The kit is for periodic verification of the sensor, and calibration if required.
The verification kit includes an integrated turbiglass (solid reference). The turbiglass value is
determined by the user for each turbidity InPro 86X0 i e series sensor the kit is used with (the
v erification kit is supplied without a sensor). It is highly recommended to use the verification kit
to determine the turbiglass value, prior to installing a new sensor.
Procedure
Notice:
It is possible that matching values may differ slightly between two new sensors. This does not
mean a sensor or the verification kit is at fault. Please consult your local METTLER TOLEDO
r epresentative for further information.
Turbidity verification with turbiglass
Filler hose
Varivent clamp
Transmitter cable
InPro 86X0 i e Series
30 714 692 A
The verification kit has to be matched with each InPro 86X0 i e series sensor before
it can be used for verification. It is necessary to record the matching value(s) for
future reference. Each subsequent verification of a sensor is based on these re-
corded values.
Discharge tube
Verification kit with
turbiglass
Sensor
Firm table