Methods

GB

Aluminium with Tablet

AL
0.01 – 0.3 mg/l
Tablet
Powder
Fill a clean vial (24 mm Ø) with 10 ml of the water sample and
0.0.0
perform zero calibration (see "Operation").
Tablet
Powder
Add one ALUMINIUM No. 1 tablet straight from the foil to the 10 ml
water sample and crush the tablet using a clean stirring rod (dissolve
the tablet).
Add one ALUMINIUM No. 2 tablet straight from the foil to the same
water sample and crush the tablet using a clean stirring rod.
Close the vial tightly with the cap and swirl gently several times until
the tablet is dissolved.
Place the vial in the sample chamber making sure that the
are aligned.
Wait for a reaction period of 5 minutes.
Zero
!
Test
(Countdown can be activated, see page 25)
AL
The method symbol flashes for approx. 3 seconds.
Tablet
Powder
RESULT
The result is shown in the display in mg/l Aluminium.
Tablet
Powder
Tolerance: ± 0.05 mg/l Al