6.2.1.10 Internal Fault

6.2.1.11 System is Cooling
6.2.1.12 Cooling Shutdown
A notice to change the illuminator dust and particulate filter is displayed every 250 hours. It is advised to
change the filter every 250 hours to maintain thermal efficiency. To remove the message, enter the menu
and press the reset filter icon. See section 6.2.2.4.1.
LIT-219 Sunoptic Surgical
Rev. F
A fault is detected in the internal control system.
ACTION: Restart the illuminator.
If the fault persists, the unit needs servicing.
Displays when the lamp module becomes overheated.
The lamp will turn OFF. If acceptable cooling is achieved
within ten seconds, the lamp will reignite and resume
normal operation.
If lamp module is unable to achieve adequate cooling, the
lamp will turn OFF.
ACTION: Disconnect unit from AC power. Check air filter
for clogging. Replace if necessary. Check vents for
obstruction.
If fault persists, the unit needs servicing.
®