6.2.1.6
Fan Fault
6.2.1.7
Dimmer Calibration Error
6.2.1.8
Lamp Fault
6.2.1.9
Lamp Communication Fault
LIT-218 Sunoptic Surgical
Rev. F
In the event of a fan fault, the device will turn off the lamp
and display this screen.
ACTION: Restart the illuminator.
If fan fault persists, the unit will need servicing.
On startup, the system calibrates the dimmer plate position.
If this process fails, this screen will display, and the lamp
will not ignite.
ACTION: Restart the illuminator.
If the fault persists, the unit needs servicing.
Signifies that the lamp will not ignite or stay lit.
ACTION: Replace lamp module.
If the fault persists, the unit needs servicing.
This screen is displayed when the control system cannot
communicate with the lamp module.
ACTION: Restart the illuminator.
If the fault persists, the unit needs servicing
®
Page 12 of 146
(English)