WARNINGS
This symbol on the instrument indicates that the user should refer to the
manual for operating instructions.
•
Shipping boxes containing an Analyzer may weigh over 40 lb —
necessitating a two-person lift technique when handling. Such boxes
are labeled with this warning, supporting OSHA standards intended to
protect employees from injury.
•
If this instrument is used in a manner not specified by SUEZ, the safety
protection provided by the instrument may be impaired.
•
To avoid injury and to ensure accurate analysis results, ensure that the
Analyzer door (On-Line TOC Analyzer) is closed and latched or side
panels (Laboratory and Portable TOC Analyzers) are on and in place
during normal operation.
•
The Analyzer requires an A/C mains supply of 100-240 volts and requires
the correct power cord for safe operation. Use only the supplied power
cord with the Analyzer, which meets this A/C mains supply requirement.
•
For safety reasons and to avoid the possibility of corrupting the
database, ALWAYS follow this procedure when shutting down the
Analyzer. First stop analysis AND stop the operating system (using the
Shut down button in the Analyzer's firmware) BEFORE pressing the
Analyzer's power switch to Off or unplugging the power cord,
•
Any operation requiring access to the inside of the Analyzer, including
installation of maintenance items, could result in injury. To avoid
potentially dangerous electrical shock, take the following precautions.
BEFORE opening the Analyzer, stop any analysis, shut down the
operating system (using the Shut down button in the Analyzer's
firmware) and turn off power and disconnect from the power supply.
•
Electrical conduit and output and alarm connections should be installed
by a qualified electrician. ESD protection should also be exercised.
•
This is a Class A product. In a domestic environment, this product
may cause electromagnetic interference. Take adequate measures to
correct the interference.
•
Hazardous reagents (ammonium persulfate and phosphoric acid) are
used in the Analyzer. Before installing a reagent, read the corresponding
Material Safety Data Sheet (MSDS) for proper handling precautions
and spill or leak procedures. The MSDS are contained in the pouch
located on the outside of the reagent shipping box.
•
Installation of reagents and most Analyzer maintenance requires access
to the inside of the Analyzer. To avoid potentially dangerous shock do
this. BEFORE opening the Analyzer, stop any analysis and shut down
the operating system (using the Shut Down button in the Analyzer's
firmware), then turn off power and disconnect from the power supply.
•
To avoid exposure to the chemical reagents, wear acid-resistant gloves
and safety goggles.
•
Reagent cartridges are for single-use only. Do NOT refill. Any refilling or
reusing of reagent cartridges will void all Analyzer and parts warranties
and nullify any performance claims.
•
After removing the plug on the BLOW MOLD type reagent cartridge,
keep the cartridge in an upright position before attaching and detaching
the reagent supply line to keep chemical from dripping out and avoid
exposure to the chemical reagent. Save the plug after removing to
later reinsert into the nozzle inlet after removing the cartridge from the
Analyzer during a cartridge replacement process.
•
To protect against accidental exposure to ultra-violet radiation, do not
operate the UV lamp outside of its protective housing.
•
(On-Line and Portable TOC Analyzers) Water in the iOS System may
be hot. Before inserting a vial into the iOS System after operating in On-
Line mode, slide the door open and wait 30 seconds to allow sample to
completely drain. Inserting a vial before draining can result in hot water
spray projecting upward out of the iOS System.
•
(Laboratory and Portable TOC Analyzers) For continued protection
against fire hazard, replace fuse with the same type and rating.
(On-Line TOC Analyzer) This symbol indicates the protective earth
terminal (ground) for the Analyzer.
The iOS System and vial ports contain sharp needles designed to
pierce the septa of sample vials. Do NOT put fingers or inappropriate
materials into the iOS System or vial port.
2018 SUEZ. All rights reserved.
20
When servicing parts inside the Analyzer, ensure that power to
the Analyzer is OFF, and keep hands clear of the reagent syringe
assemblies. The syringes are controlled by moving parts that can pinch
skin.
HARZARDOUS WASTE DISPOSAL
Hazardous reagents (ammonium persulfate and phosphoric acid) are
used in the Analyzer. The waste stream from the instrument is acidic
and must be disposed of properly. Consult your federal, state, and local
government regulations.
•
The UV lamp and the display screen contain mercury and may be
considered hazardous material in your local area. Dispose of these
items in accordance with federal, state, or local government regulations.
•
In the case of any broken or damaged UV lamp, handle the remains in
accordance with your organization's toxic waste handling procedure and
dispose of them in accordance with federal, state, or local government
regulations.
This symbol indicates the European Union Directive 2002/96/EC for
waste electrical and electronic equipment (WEEE) requires disposing
the Analyzer separately from standard waste.
OPERATIONAL CAUTIONS
•
Ensure that the DI water Reservoir is full, particularly when running
samples with high TOC or high salt concentrations. After running high
TOC or salt samples, always rinse the Analyzer by running low-TOC DI
water through a sample cycle.
•
To avoid false TOC readings and possible damage to the Analyzer,
always ensure that the sample inlet is open and the DI water reservoir
is filled before starting analysis.
•
Operation of the Analyzer without the in-line filter on the sample inlet
line will damage the Analyzer and void the warranty. To avoid damaging
the Analyzer, install the filter and replace the filter element, as needed.
•
To avoid damage to the DI water pump, ensure that the DI water
reservoir is filled BEFORE turning On the Analyzer.
•
To avoid false TOC readings and possible damage to the Analyzer,
always ensure that the sample is flowing through the iOS System and
the DI water reservoir is filled before starting analysis.
•
For performance within specifications on ozonated water systems,
an Ozone Destruct Kit must be purchased from SUEZ and installed
according to instructions.
•
When cleaning the Analyzer, always apply liquid to the cloth directly.
This symbol on the instrument indicates that the product does contain
restricted substances included in China RoHS II. Please refer to the
following table.
Note:
This product contains a UV Lamp (with Hg) which meets the
exemption defined in Annex III 4(a) of RoHS Directive 2011/65/EC.