Descargar Imprimir esta página

Invacare HomeFill II Manual De Instrucciones página 25

Ocultar thumbs Ver también para HomeFill II:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 147
PROBLEM
Compressor cycles On and Off
or
O
Below Normal (yellow) and Filling
2
(green) lights are alternating
or
O
Below Normal(yellow) light goes On
2
and Off
O
Below Normal (YELLOW) light stays
2
On.
1145804-E
All manuals and user guides at all-guides.com
SOLUTION
2. Check that the cylinder is connected properly. Turn the compressor Off (O). Remove the
cylinder and reinstall. Turn the compressor On ( ).
3. If cylinder is not full and is connected properly, the compressor may have internal failure.
Contact your home health care provider or Invacare.
1. Check the concentrator liter flow setting is at 2.5 L/min. or less. If necessary, adjust the liter flow
setting (refer to the concentrator user manual). Wait at least 20 minutes for the GREEN light to
appear. If it does not appear, proceed to step 2.
2. The concentrator may need service, contact your home health care provider or Invacare.
1. Ensure the concentrator has warmed up for at least 20 minutes. If it has not been 20 minutes,
turn the compressor Off (O) until the concentrator warms up. If the YELLOW light still remains
on proceed to step 2.
2. Inspect the connection between the compressor and the concentrator for damage or pinching. If
damaged, replace. If pinched, straighten tubing. Turn the compressor Off (O) for at least 10 seconds.
Turn the compressor On ( ). If the YELLOW light still remains on after 10 minutes, proceed to step 3.
3. The concentrator may need service, contact your home health care provider or Invacare.
Troubleshooting
25

Publicidad

loading

Este manual también es adecuado para:

Ioh200aw