NOTICE:
•
Changes or modifications to the device will nullify the
user warranty.
•
Do not drop or insert any object into any opening. This
may damage the unit.
•
Do not press the buttons with excessive force or with
pointed objects.
To keep your digital blood pressure monitor in the best con-
dition and protect the unit from damage follow the direc-
tions listed below:
• Clean your device and cuff carefully only with a slightly
moistened soft cloth and dry it immediately with a soft
dry cloth. Do not press.
• When storing the device, make sure that no heavy objects
are placed on top of it.
• If the unit is stored near freezing, allow it to acclimate to
room temperature before use.
• Always store the unit in the box after use.
• This unit may not meet its performance specifications if
stored or used outside of the temperature and humidity
ranges (see section "10. Technical Specifications").
• Use the unit consistent with the instruction provided in this
manual.
All manuals and user guides at all-guides.com
8. Error messages /troubleshooting
Error Display/
Symptoms
Er1 or Er2
Er3 or Er4
Er5
Er6
Er0, Er7, Er8,
or ErA
Unit does not
turn on when
the
is pushed.
Note: If the unit still does not work, contact customer ser-
vice. Please see warranty for service contact.
18
Possible Causes
The blood-pressure
values are unusually
high or low.
The cuff is too tight
or too loose.
The inflation pres-
sure is higher than
300 mmHg.
Inflation takes lon-
ger than 3 minutes.
There is a system or
device error.
Batteries have run
down.
button
Battery polarities
have been posi-
tioned incorrectly.
The unit is defec-
tive.
The batteries are
almost empty.
Correction
Repeat the measurement.
Make sure that the cuff
is positioned correctly,
and that you do not move
or talk.
Re-insert the batteries if
necessary, or else replace
them.
Contact customer service.
See warranty section.
Replace the batteries.
Re-insert the batteries in
the correct position.
Contact customer service.
See warranty section.
Replace the batteries.