div 01
LF2
LINEAL CALIBRATION
Up to six steps of calibration W0 ~W6
Make sure that the plate of the scale is empty before starting the calibration.
Press the key
The previous points of calibration will be removed and the message CLEAR is going to appear on the screen.
Press the key
Press the key
Press the key
Press the key
DISPLAY
LF3
W0
W1
W2
W3
W4
LF3
40
• Use
and then
and go back to the menu LF2).
• Use
,
and
from the menu and the scale will begin again automatically).
LF 3
or the first point of calibration.
to proceed with the next point of calibration.
to go back to the previous point of calibration.
to save everything.
to finish the calibration and go back to the menu
•Press the Key
to start or the key
matically.
•Press the Key
to calibrate the zero.
( press the key
to exit from the calibration and go back to the menu LF3)
• Place 1/3 of the maximum weight on the plate and press
(press the key
to exit from the calibration and go back to the menu LF3).
•Place 2/3 of the maximum weight on the plate and press
(press the key
to exit from the calibration and go back to the menu LF3).
•Place the maximum weight on the plate and press the key
the key
to exit from the calibration and go back to the menu LF3).
•Press the key
to complete the lineal calibration. (press the key
bration and go back to the menu LF3)
•Use the keys
,
and then the key
key
to exit from the menu and the scale will begin again automatically.
tho change the step.(Press the key
To continue with the other configuration or press the key
LF3.
DESCRIPTION AND SEQUENCE OF USE
to exit the menu and the scale will start again auto-
to continue with other adjustments and press the
USER'S MANUAL ABD
to exit from the configuration
to calibrate 2/3 th the capacity.
to calibrate 2/3 th the capacity.
to calibrate the full capacity. (press
to exit from the cali-
EN
EN
to exit