Assemble the stand pole RV 3.1 (Accessories)
• Assemble the condenser locking device according to the
diagram.
• Mount the plate (1) using the thumb screw (2).
• Put the support rod (4) on the plate (1) and attach it with
the screw nut (3).
• Attach the rubber protector (5).
• Fasten the Velcro
band (6) to the support rod (4).
®
• Secure the vertical glassware with the Velcro
3
The device is maintenance-free. It is only subject to the
natural wear and tear of components and their statistical
failure rate.
The seal on the glass condenser should be checked at regu-
lar intervals and replaced if necessary.
Cleaning
Remove the device from the
mains before cleaning.
Only use cleaning materials recommended by IKA
Dirt
Cleaning agent
Dyes
Isopropyl alcohol
Building materials Water containing detergent/isopropyl
alcohol
Cosmetics
Water containing detergent/isopropyl
alcohol
Food
Water containing detergent
Fuels
Water containing detergent
Other materials
Please consult IKA
Wear protective gloves during cleaning the devices.
Electrical devices may not be placed in the cleansing agent
for the purpose of cleaning.
No stand is required if
the condenser has been
properly fitted and secured
by tightening the condenser
union nut on the rotary drive.
The purpose of the stand
is solely to prevent the
band (6).
condenser from twisting.
®
6
CAUTION
If
the
incorrectly,
5
forces at the Velcro strap
can cause stresses within
the glass, which in turn can
cause damage to the glass
condenser.
4
After the condenser has been
fitted, the stand is attached
to the rotary drive. Make sure
that the condenser is installed
parallelly to the lift body.
2
1
Maintenance and cleaning
Do not allow moisture to get into the device when clean-
ing.
Before using another than the recommended method for
cleaning or decontamination, the user must ascertain with
IKA
that this method does not destroy the device.
®
Ordering spare parts
When ordering spare parts, please give:
- Device type,
- Serial number, see rating plate,
®
:
- Position number and description of spare part, see
www.ika.com.
Repairs
Please only send devices in for repair that have been
cleaned and are free of materials which might present
health hazards.
For this, use the "Decontamination Certificate" form
which you can obtain from IKA
sion for printing from the IKA
If your device requires repair, return it in its original packag-
®
ing. Storage packaging is not sufficient when sending the
device - also use appropriate transport packaging.
stand
is
fitted
the
strong
or can download a ver-
®
®
website at www.ika.com.