Descargar Imprimir esta página

Taking A Blood Pressure Measurement; Manual Inflation; How Do I Evaluate My Blood Pressure; Appearance Of The Irregular Heartbeat (Ihb) Symbol - Microlife BP B2 Easy Manual De Instrucciones

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 10

3. Taking a blood pressure measurement

1. Press the ON/OFF button 1 to start the measurement.
2. The cuff will now pump up automatically. Relax, do not move
and do not tense your arm muscles until the measurement
result is displayed. Breathe normally and do not talk.
3. The cuff fit check AR on the display indicates that the cuff is
perfectly placed. If the icon AR-A appears, the cuff is fitted
suboptimally, but it is still ok to measure.
4. When the correct pressure is reached, the pumping stops and
the pressure falls gradually. If the required pressure was not
reached, the device will automatically pump some more air into
the cuff.
5. During the measurement, the pulse indicator AQ flashes in the
display.
6. The result, comprising the systolic 9 and the diastolic AT blood
pressure and the pulse rate AK are displayed. Note also the
explanations on further display symbols in this booklet.
7. When the device has finished measuring, remove the cuff.
8. Switch off the device. (The monitor does switch off automati-
cally after approx. 1 min.).
You can stop the measurement at any time by pressing the
ON/OFF button (e.g. if you feel uneasy or an unpleasant
pressure sensation).
This monitor is specially tested for use in pregnancy and
pre-eclampsia. When you detect unusual high readings in
pregnancy, you should measure after a short while again
(eg. 1 hour). If the reading is still too high, consult your
doctor or gynecologist.

Manual inflation

In case of high systolic blood pressure (e.g. above 135
mmHg), it can be an advantage to set the pressure individually.
Press the ON/OFF button after the monitor has been pumped up
to a level of approx. 30 mmHg (shown on the display). Keep the
button pressed until the pressure is about 40 mmHg above the
expected systolic value – then release the button.

How do I evaluate my blood pressure

The triangle on the left-hand edge of the display AO points at the
range within which the measured blood pressure value lies. The
value is either within the optimum (white), elevated (hatched gray)
or high (black) range. The classification corresponds to the
following ranges defined by international guidelines (ESH, ESC,
JSH). Data in mmHg.
Microlife BP B2 Easy
Range
Systolic Diastolic Recommendation
≥135
1. blood pressure
too high
130 - 134 80 - 84
2. blood pressure
elevated
3. blood pressure
<130
normal
The higher value is the one that determines the evaluation.
Example: a blood pressure value of 140/80 mmHg or a value of
130/90 mmHg indicates «blood pressure too high».

Appearance of the irregular heartbeat (IHB) symbol

This symbol AN indicates that an irregular heartbeat was detected.
In this case, the measured blood pressure may deviate from your
actual blood pressure values. It is recommended to repeat the
measurement.
Information for the doctor in case of repeated appearance of
the IHB symbol:
This device is an oscillometric blood pressure monitor that also
measures the pulse during blood pressure measurement and
indicates when the heart rate is irregular.

4. Data memory

This device always stores the last result at the end of the measure-
ment. To recall the reading, press and hold the ON/OFF button 1
while the device is switched off. Release the button as soon as all
display elements are shown. The stored result will then be shown,
indicated with the letter «M» AP.

5. Battery indicator and battery change

Low battery

When the batteries are approximately ¾ empty the battery symbol
AL will flash as soon as the device is switched on (partly filled
battery displayed). Although the device will continue to measure
reliably, you should obtain replacement batteries.

Flat battery - replacement

When the batteries are flat, the battery symbol AL will flash as soon
as the device is switched on (flat battery displayed). You cannot
take any further measurements and must replace the batteries.
≥85
Seek medical
advice
Self-check
<80
Self-check
3
EN

Publicidad

loading