EN-4
Rotors must be loaded symmetrically and with equal weight. The adapter may only be loaded with the
compatible containers. The weight differences between the filled vessels should be kept as low as possible.
Therefore we recommend weighing them with a balance.
2.3 Loading samples for swing out rotors
Swing out rotors should only be used if all positions are occupied with buckets and racks, and the
diagonally opposing pair must be buckets and racks of the same size and type. Do not mix different types
of buckets and racks diagonally! It is allowed to operate e.g. a 4-place-rotor with 2 loaded buckets only but
the loaded buckets must be opposite to each other. Make sure that the unloaded racks also be put inside
the buckets. The bolts on the rotor must be greased with the "High TEF oil". The sample tubes have to be
filled evenly by eye and put into the tube racks.
Correct operation
3. OPERATION
3.1 Display Panel
TM
Frontier
2000 Centrifuges
Correct operation
Wrong operation
Wrong operation